H601-K-FRAME-GLASS-RECESSED
The Glass T Plate creates a warm and optically transparent surface perfectly recessed relative to the stage. It employs Indium Tin Oxide (ITO) coating to obtain uniform heating throughout the glass surface and a surface sensor to ensure the highest accuracy on temperature reading.
H601-K-FRAME-GLASS-RECESSED features anti-scratch, anti-fingerprint, chemically strengthened Dragontrail glass cased in rigid aluminum frame for the highest resistance against breakage and scratches.
Protected by 10 year warranty on glass damage.
It requires H401-T-CONTROLLER or H401-T-PENNY.
Features
- Fits stages with k-frame insert (160x110mm), compatible with the following XY stage: Marzhauser (all models), Ludl (all models), ASI (all models, including z-piezo), Prior H107 and H117, Leica (k-frame), Zeiss (k-frame), Olympus (k-frame)
- Recessed with respect to the rest of the XY stage. Rises the focal plane by only 1.5 mm.
- Temperature range: from ambient to 60°C
- Accuracy on specimen: ±0.3°C
- ITO glass coating ensures uniform glass heating
- Glass thickness: 0.5 mm
- Aluminum frame improves rigidity
- Anti-scratch, anti-fingerprint, chemically strengthened Dragontrail glass
- 10 year warranty on glass damage
H401-T-CONTROLLER
2-Channel Temperature Controller with 4.3” touch screen interface compatible with all H401 heating devices. One controller can control independently up to two items.
External temperature sensor (included) and self calibration routines allows easy calibration of the heating devices for the highest accuracy on sample temperature.
On board data logging and download procedures to USB drives allow to record temperature as a function of time and to log calibration events and alarms.
The controller is equipped with password protected routines to re-calibrate the sensors against certified thermometers.
Features
- T – range: from ambient to 60°C
- T – accuracy on specimen: ± 0.3°C
- 4.3” touch screen interface
- External T – sensor (included)
- On-board Data logging and download routines to USB drive
- On-board logging of calibration and alarm events
- Alarm buzzer and External Alarm Connector
- One controllers can control independently up to two devices
- Self calibration routines allow to improve temperature accuracy
- Password protected procedure for periodic calibration against certified thermometer
H401-T-PENNY
H401-T-PENNY is a compact single-channel Temperature Controller compatible with any OKOLAB heating device that combines performance and simplicity at an unbeatable price.
Features
- Temperature range: from ambient to 60°C
- Set point resolution: 0.1°C
- Calibration via addition of offset
- Maximum Output: 60 W
- Dimensions: 83.65×45.05×20.20 mm
- Weight: 47.50 g